I.USE
Used for determining the flash and fire point of petroleum products. It consists of a brass cup mounted on an electric heater
with temperature controller.
1. This instrument is designed as per GB/T261 and ASTM D93.
2. The heating power is continuously adjustable. The power is shown visually by voltmeter.The temperature control mode is advanced and reasonable.
3. The structure is designed small and exquisite.The stainless-steel table board is beautiful and easy to clean.
4. Operation is easy.Test results are precise.It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes
II. Technical Specification
1. Power supply: AC(220V±10%)V, 50HZ
2. Heating set:
(1) The furnace body is silicon carbide material, and the power is 600w electric heating wire.
(2) heating power: (0-600) w, adjustable
3. Sample Heating rate: (1~12)ºC/min, manual control
4. Electromagnetic stirring device:
(1) Stirring motor: 45TYZ Constant speed motor
(2) Transmission mode: Flexible shaft coupling
(3) Stirring blade size: 8mm X 40mm
5. Stirring speed:
(1) (90~120) rotation/min, Suitable for experimental step A.
(2) (250±10) rotation/min, Suitable for experimental step B.
6. Standard oil cup:
(1) inner diameter: 50.7~50.8 mm
(2) Depth: 55.7~56.0mm
(3) Test oil volume depth: 33.9~34.3mm
(4) Test oil capacity: about 70ml
7. Fire device
(1) Fire source: Coalgas
(2) Fire igniter radius: (0.7~0.8)mm
8. Thermometer: Bar mercury thermometer
(1) Temperature range (-5~110)ºC, accuracy: 0.5ºC
(2) Temperature range (20~150)ºC, accuracy: 1.0ºC
(3) Temperature range (90~370)ºC, accuracy: 2.0ºC
9. Ambient temperature: ≤35ºC
10. Relative humidity: ≤85%
11. Machine power consumption: not more than 650W
12. Division: 370x320x300mm